Main Shortcut

Equipment support

KORAM
Korea Institute for Rare Metals
Start of the main

Equipment support

Specific Surface Area Analyzer(BET)

Specific Surface Area Analyzer(BET) Image

Equipment information
Field Analysis
Manufacturer Microtrac
Model Adsotrac DN-04
Related Sites
Management personnel 서석준
Responsible for contacts 032-226-1358
Fee 40,000₩/Sample

- Principle : Nitrogen adsorption method
- Function 
    BET surface area measurement(Single and Multi-point method)
    Direct reference comparison
    Langmuir surface area measurement
- Method : Dynamic Gas Flow method
- Working gas : N2(adsorbate), He(Career) 99.99%
- Nitrogen patial pressure : 0.05~0.35
- Range : (0.01m2/g≤Surface area; no up limit)
- Repeatability : ≤ ±2%
- Measurement Time 
    BET method for each sample < 25 min.(1point)
    Reference comparison method 8 min.
    Dimension : 500(W)350(D)590(H)
- Weight(kg) : 32
- Adsorption gas: high purity nitrogen.  Ar, Kr, CO2, etc. available 
- Nitrogen partial pressure:P/Po5×10-7----0.995
- Pressure Accuracy:   ±0.15%  
- Measuring range: Surface area ≥ 0.01M2/g , above;  Pore size  0.35 ~ 400 nm
- Repeatability:≤ ±2 %
- Test Efficiency:Multi-point BET surface area per sample: 10 ~ 15 minutes; 
    Absorption and desorption isotherms measures pore size distribution: 3 ~ 5h; 
   (Test efficiency exponential growth with the No of analysis stations)      
- Operation Mode: Automatic
- Measuretime : BET method for each sample  < 25min (1 point)
    Reference comparison method  8min
- BET Dimension(W*D*H;mm) : 600×450×680, 38kg
- Vacumm heating system(W*D*H;mm) : 400×350×550, 10kg