Electron beam Melting Furnace

◈ Configuration and performance.
1) VACUUM CHAMBER
MAIN CHAMBER is a STAINLESS STEEL circular chamber in which the ARC GUN is installed inside the CHAMBER, and a water-cooled HEARTH is attached to the lower chamber to detach it in a UP, DOWN manner.
A see-through window should be installed in the area where monitoring is required to allow observation of the interior during work, and there should be no damage caused by ARC heat.
① Type: Round vertical upper and lower double chambers
② SIZE(mm): 800800 X Height H-1200 (can be changed after consultation)
③ Material: SUS304.
④ Install the front of VIEW PORT.
⑤ E-BEAM attachment PORT.
⑥ POWER FEEDTHROUGH PORT
⑦ COLD CRUCIBLE HEARTH ELEVATION PORT
⑧ Material supply feeding device attachment port
⑨ VACUUM PUMPING PORT
2) Vacuum exhaust system.
It is designed to operate the ROTARY PUMP and vacuum valve automatically or manually,
After the operation of the ROTARY PUMP, the degree of vacuum will be 10-2 Torr or higher (within 15 minutes).
① OIL ROTARY PUMP
⑴ Pumping speed: 1000 L/min.
⑵ Reach vacuum: 5X10-3Torr (pump self-performance)
⑶ 220V, 3P, 60HZ
⑷MAKER : KODIVAC
② VALVE
⑴ MAIN VALVE 10-INCH (PNEUMATIC)
⑵ ROUGHING VALVE 2-INCH (PNEUMATIC)
⑶ MAKER : YEIN-TECH
③ DIFFUSION PUMP
⑴ 2000 L/SEC
⑵ WATER COOLING TYPE
⑶ Including CONTROLLER UNIT-GAUGE.
⑷ MAKER : KODIVAC
④ HEPA filter to remove fume and dust.
3) Quantitative discharge feeder.
① Quantitative control controller.
② Vibration amplitude sensor attached.
③ Slope Chute attached.
④ Dustproof pads.
⑤ Input power: 220V 3P
4) MIAN FRAME
① Welded steel structure.
② SIZE : 2500*2500*2500H
③ Material: H-shaped steel, each structural tube, etc.
④ Stairway to connect the 2nd floor: Check plate applied.
⑤ Post-treatment: Paint
5) MAIN CONTROL PANEL (MANUAL TYPE)
① VACUUM CONTROL : MANUAL TYPE
② GAS VALVE ON,OFF SWITCH
④ HEARTH ELEVATION CONTROL
⑴ Form : MANUAL SWITCH ON, OFF UP, DOWN (LIMIT-OFF)
⑤ INTER LOCK SYSTEM
⑴ Designed to allow the equipment to pause and return in case of abnormal operation.
⑵ Coolant pressure drop mark (lamp on, buzzer on)
⑶ Lack of compressed air pressure (lamp on sub-scale)
⑥ VACUUM GAUGE